| System |
Automated, discrete clinical chemistry
analyzer intended for the in vitro quantitative/qualitative
determination of analytes in body fluids
|
 |
 |
 |
| System Components |
Self contained floor model analyzer comprised
of the analytical unit and control unit
|
 |
 |
 |
| Sample Throughput |
300 photometric tests per hour
|
 |
 |
 |
| Number of Channels |
42 channels and 3 ISE's
|
 |
 |
 |
| Programmable Parameters |
Maximum: 117 photometric, 3 ISE tests, 8
formulas, 3 serum indices
|
 |
 |
 |
| Sample Material |
Serum/plasma, urine, CSF
|
 |
 |
 |
| Sample Input/Output |
Sample disk with 108 positions plus: W1
position for ISE cleaning solution, W2 position for ISE activator,
Real STAT interrupt
|
 |
 |
 |
| Sample Cups and Tubes |
|
Sample Container
|
Diameter x length
|
|
Primary sample tube
|
16mm x 100mm
|
|
Primary sample tube
|
16mm x 75mm
|
|
Primary sample tube
|
13mm x 100mm
|
|
Primary sample tube
|
13mm x 75mm
|
|
Hitachi standard cup
|
17mm x 38mm, 2.5ml
|
|
Hitachi micro cup
|
8mm x 37mm, 1.5ml
|
|
False bottom tube
|
13mm x 75mm, one type definable
|
|
 |
 |
 |
| Sample Clot Detection |
Available (pressure sesitive clot detection
system)
|
 |
 |
 |
| Liquid Level Sensor |
Capacitance sensing technology
|
 |
 |
 |
| Sample Barcode Types |
NW7 (Codabar); Code 39; ITF; Code 128
|
 |
 |
 |
| Control Unit |
Microsoft Windows XP, Pentium IV Processor
|
 |
 |
 |
| System Interfaces |
RS–232 serial interface
|
 |
 |
 |
| Sample Data Base |
10,000 samples (routine, STAT, and QC samples)
|
 |
 |
 |
| Calibration Methods |
Start–up, re–calibration. For
photometric assays: Linear, non–linear multi points, 2 point
calibration, K–factor. Up to 100 calibrators
pre–programmable. Storage of up to 180 curves. Preventative
calibration of stand–by cassettes. Two k–factor can be
defined for different sample types.
|
 |
 |
 |
| QC Methods |
Real–time QC, individual QC, cummulative
QC. Up to 100 different controls pre–programmable.
Preventative QC after calibration of stand–by cassettes.
Auto QC (without operator intervention).
|
 |
 |
 |
| Electrical
Requirements |
|
Power rating
|
AC 208V AC/60Hz (US/Canada)
|
|
Power supply fluctuation
|
No significant power supply fluctuation
(operating on 208/230V AC, max power supply change: ± 10%)
|
|
Overvoltage category
|
II
|
|
Polution degree
|
2
|
|
Power consumption
|
1.5 kVA for analytical unit; 0.5 kVA for
control unit
|
|
Electrical installation
|
Technical standard class C,
Required eathing < 10 Ω,
Bonding impedance < 0.1Ω at 30 A,
Insulation resistance > 10MΩ at 500 V
|
|
 |
 |
 |
| Water/Waste Requirements |
|
Bacteria–free, deionized water
|
< 10 cfu/mL
|
|
Conductivity
|
> 10 MΩ resistance
|
|
Water pressure
|
5015 to 25 psi (1.054 – 1.76g/cm2)
|
|
Water supply volume
|
40 L/h
|
|
 |
 |
 |
| Operating Conditions |
|
Ambient temperature
|
18 to 32° C
|
|
Ambient humidity
|
45 – 85% (non–condensing)
|
|
Noise Output
|
<70 dB(A) for surrounding
|
|
 |
 |
 |
| Physical Dimensions |
|
Depth
|
859 mm
|
34 in
|
|
Height
|
1260 mm
|
50 in
|
|
Height level of monitor
|
1380–1570
mm
|
54.3–61.8 in
|
|
Width
|
1325 mm
|
52 in
|
|
Weight
|
270 kg
|
595 lb
|
|
 |
 |
 |
| Sampling System |
|
Sample pipetting volume
|
1.5–35 µL in 0.1 µL
increments
|
|
Detection of sample clogging
|
Pressure sensitive clot detection system
|
|
Liquid Level Sensor
|
Capacitance sensing technology
|
|
 |
 |
 |